· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots distributed from March 1, 2022. | Centrella Pro+ Surface | P7923A04 P7923A03 |
| All produced after June 9, 2020. | Pro+ MRS Surface For VersaCare With X-Ray | P3255A01 |
| All produced after June 9, 2020. | Pro+ MRS Surface With X-Ray | P7924A02 P7924A01 |
| All produced after June 9, 2020. | Centrella Max Surface | P7922A01 P7922A02 |
| All produced after June 9, 2020. | Centrella Pro+ Surface With X-Ray | P7923A02 P7923A01 |
| All lots distributed from March 1, 2022. | Pro+ MRS Surface For VersaCare | P3255A02 |
| All lots distributed from March 1, 2022. | Pro+ MRS Surface | P7924A03 P7924A04 |
| All lots distributed from March 1, 2022. | Pro+ MRS Surface For Hillrom 900 With X-Ray | P006800A01 |
Baxter Corporation is issuing a correction for the Centrella Max and Pro+ Mattress Replacement System (MRS) Hospital Bed Mattresses due to the potential for the top cover to delaminate. Contributing causes of delamination include improper cleaning of the product (i.e., excessive use of cleaning agents containing bleach) and a lack of adequate details on the appropriate use of cleaning agents containing bleach in the product instructions for use (IFU).
Recall Start Date: May 13, 2024