· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots. | Proclaim 7 Implantable Pulse Generator | 3667 3663 |
| All lots. | Proclaim DRG Implantable Pulse Generator | 3664 |
| All lots. | Proclaim XR 5 Implantable Pulse Generator | 3660 |
| All lots. | Proclaim XR 7 Implantable Pulse Generator | 3662 |
| All lots. | Proclaim 5 Implantable Pulse Generator | 3665 3661 |
Abbott has initiated a voluntary medical device recall of Implantable Pulse Generator (IPG) for patients with non-rechargeable proclaim neurostimulation systems as the duration between the IPG reaching battery elective replacement indicator (ERI) threshold and end of service (EOS) may be 45-55% shorter than indicated in the product labeling.
Recall Start Date: June 6, 2024