CALiaGold PierceTube

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: CALiaGold PierceTube
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
40045
31390
40037
CALiaGold PierceTube1151300

Issue

Following a customer complaint, Sentinel detected a defect in three (3) lots of the collection tube CALiaGold PierceTube, ref. 1151300: lot 31390, 40037, 40045. The defect is an inner plastic crack that connects the two chambers of the collection tube, causing a reduction of the buffer volume needed for the measurement. It cannot be excluded that a significant volume decrease may lead to the impossibility to perform the test. The root cause of the plastic crack is still under investigation.