Health Products

Alinity HQ Analyzer

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Haematology
Product: Alinity HQ Analyzer
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Alinity HQ Analyzer09P68-01

Issue

Abbott has identified the following issues when using Alinity HQ Analyzer software version 5.6 and below which will be corrected in software version 5.9.

Issue 1: When processing a CBC+diff sample that immediately follows a CBC+diff+retic sample, diluent may backflow into the RBC dilution cup. This may result in falsely low red blood cell (RBC) results and consequently falsely high mean cell hemoglobin concentration (MCHC), falsely high mean cell hemoglobin (MCH), and falsely low hematocrit (HCT).

Issue 2: An overestimation of the basophil (baso) count may occur on some samples when cell events are incorrectly counted as baso during sample analysis. this may result in falsely increased baso counts and %b (%baso).

Recall start date: June 21, 2024