Health Products

Luminos dRF Max, MOBILETT Elara Max, YSIO Max and YSIO X.Pree

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Luminos dRF Max, MOBILETT Elara Max, YSIO Max and YSIO X.Pree
Issue: Medical devices - Device compatibility
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
11107444MOBILETT Elara MaxVF10K
11107464YSIO X.PreeVA10F_HF03
10762470YSIO MaxVF10K_SP05
VF10K
VF11H
VE10ZA
VF10F_HF05
10762471Luminos dRF MaxVF10K_SP05

Issue

Siemens Healthineers was informed by the detector manufacturer that magnetic connectors may have been supplied as spare parts for portable detectors. If magnetic connection units are installed, a permanent (static) magnetic field is generated that exists independently of the operation of the detector. This can affect the function of implantable electronic devices such as pacemakers or implantable defibrillators if a minimum distance of >5 cm from the implanted device is not maintained.