· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| More than 10 numbers, contact manufacturer. | 5000-00-00 5000-00-00L | Arctic Sun 5000 |
| More than 10 numbers, contact manufacturer. | 6000-00-00 6000-00-00L | Arctic Sun Stat |
An internal investigation has determined that alarm 113 may not alert as expected when the device fails to reach the correct target water temperature while the device is operating in patient control mode.
Recall start date: June 27, 2024