Health Products

Arctic Sun 5000 and Arctic Sun Stat

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Arctic Sun 5000 and Arctic Sun Stat
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
More than 10 numbers, contact manufacturer.5000-00-00
5000-00-00L
Arctic Sun 5000
More than 10 numbers, contact manufacturer.6000-00-00
6000-00-00L
Arctic Sun Stat

Issue

An internal investigation has determined that alarm 113 may not alert as expected when the device fails to reach the correct target water temperature while the device is operating in patient control mode.

Recall start date: June 27, 2024