Health Products

Infinity M500 Docking Station

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Infinity M500 Docking Station
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Infinity M500 Docking StationMS20407

Issue

Draeger Medical Systems, Inc. has become aware during testing in conjunction with global market surveillance activities that the Infinity M500 Docking Station is not in full compliance to type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards.

The only ports that could be affected are: Hemo port (gray color), SpO2 port (blue color) and Temp port (white color) on M540. The specific accessories which can be connected to these ports and may be affected are the Dual Hemo MCable, SpO2 MCables, and temperature probes, respectively.

Recall Start Date: April 15, 2024