Health Products

Titan Blis-Sonic SW Scaler

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Dental
Product: Titan Blis-Sonic SW Scaler
Issue: Medical devices - Device compatibility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
4874201Titan Blis-Sonic SW Scaler261669
264563

Issue

A voluntary recall was initiated for a certain lot of Titan Scaler Tip-Perio (part number# 261669) specifically packaged with Blis-Sonic Scaler (part number# 264563) lot# 4874201 distributed between March 06, 2024 through May 10, 2024. This recall is initiated because it was discovered during production testing that the tip holder portion of the scaler tip assembly may have an oversized condition that will not allow the scaler tip to stay seated in the holder.

Recall Start Date: June 4, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime