· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| 782425 712258 743930 742049 769240 | BIOFIRE® Filmarray® Gastrointestinal (GI) Panel | RFIT-ASY-0116 RFIT-ASY-0104 |
An increased number of customer complaints associated with Vibrio/Vibrio Cholerae false positive detections by the BIOFIRE Filmarray Gastrointestinal (GI) Panel have been observed when using Remel Cary-Blair transport medium. The BIOFIRE GI panel and the Cary-Blair transport medium are performing as intended.
Per INV-19261, the root cause of the false positive Vibrio/Vibrio Cholerae results detected by the BIOFIRE GI panel is the presence of non-viable organism in the Remel Cary-Blair transport medium.
Recall Start Date: July 4, 2024