Fassier-Duval Telescopic IM System (FRT250 - 4.0-6.4 Cartridge)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Orthopaedics
Product: Fassier-Duval Telescopic IM System (FRT250 - 4.0-6.4 Cartridge)
Issue: Medical devices - Device compatibility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
F462-02
F281-04
Fassier-Duval Telescopic IM System (FRT250 - 4.0-6.4 Cartridge)FRT250

Issue

It has been identified that an incorrect sub-component, FRT240-HEX, was used in place of the FRT250-HEX during assembly. If such an FRT250 cartridge is used in surgery, it will not function correctly for the retrieval of the Fassier-Duval female component.

Recall Start Date: July 4, 2024