HeartStart Intrepid Monitor/Defibrillator

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: HeartStart Intrepid Monitor/Defibrillator
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Not applicable.Heartstart Intrepid Monitor/Defibrillator867172

Issue

When monitoring ECG using either a 5-lead or 10-lead ECG cable, the HeartStart Intrepid Monitor/Defibrillator may display intermittent ECG waveforms when the fourth limb lead is placed on the patient. The ECG tracing is normal when only three limb leads are connected (right arm, left arm, and left leg). However, when there are one or more poor ECG lead connections to the patient, the ECG tracing either displays a dashed line or is intermittent between waveform and dashed line. This failure may occur any time the HeartStart Intrepid Monitor/Defibrillator is being used to monitor a 5-lead or 12-lead ECG.

Recall Start Date: July 9, 2024