Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Trilogy EVO and EV300
Issue: Medical devices - Device compatibility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.Trilogy EVO, CanadaCA2110X12B
All lots.Trilogy EVO, O2, Canada EV300CA2200X12B

Issue

Philips Respironics has become aware that Trilogy EVO and Trilogy EV300 devices do not comply with the obstruction alarm requirements specified within ISO 80601-2-12 clause 201.12.4.108 and ISO 80601-2-72 clause 201.12.4.107.

The standards specify that the maximum delay from obstruction to alarm shall be no more than 2 breath cycles or 5 seconds, whichever is greater. The current obstruction alarm delay for these devices has been measured up to 65 seconds in certain bench test conditions. This is a delay potentially 60 seconds greater than required by the standards.

Recall start date: July 1, 2024