· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| 1900148359 1900179815 1700120620 | VITROS® XT Chemistry Product ALB-TP Slides | 684 4293 |
QuidelOrtho™ has confirmed an issue involving VITROS® XT Chemistry Products ALB-TP slides which may create dust and debris within the microslide subsystem on VITROS® XT 3400 Chemistry Systems and VITROS® XT 7600 integrated systems. Dust and debris from XT ALB-TP slides may impact VITROS® Chemistry Products Na+ slides leading to a potential increase in non-reproducible, positively, or negatively biased Na+ results.
Recall start date: July 12, 2024