Health Products

Change Healthcare Cardiology

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Change Healthcare Cardiology
Issue: Medical devices - Labelling and packaging
What to do: Customer will be notified with instructions included.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Software Version CHC14.3.2
Software Version CHC 15.0.1 (SP1)
Change Healthcare CardiologyVERSION 14.X
VERSION 15.X

Issue

Change healthcare has become aware of the use of an unapproved configuration involving misaligned regional settings between the server and client/browser. In certain workflows, this configuration can lead to incorrect date displays, potentially resulting in the month and day being incorrectly switched and affecting the accuracy of recorded dates.