Health Products

IRISpec CA/CB/CC

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: IRISpec CA/CB/CC
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
432-24 and higherIRISpec CA/CB/CC800-7211
800-7702

Issue

Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB glucose control failure, resulting in false positive results when used in conjunction with iChemVELOCITY Urine Chemistry Strips PNS 800-7204 and 800- 7212. If a QC run fails, the system will not allow the operator to run patient samples. The QC run should be repeated and pass successfully before running patient samples.

Recall start date: July 16, 2024