StealthStation S8 Application

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Neurology
Product: StealthStation S8 Application
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.StealthStation S8 Application9735762

Issue

Medtronic is writing to customers to inform them of an issue associated with the StealthStation™ S8 application, and specifically the display that shows the "distance to target" or "tip stop point" measurements. There is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation. In certain situations, this anomaly could lead to incorrect data being displayed during surgical procedures.

Recall Start Date: April 10, 2024