Health Products

Trevo Microcatheters and Synchro Devices

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular, Health products - Medical devices - Neurology
Product: Trevo Microcatheters and Synchro Devices
Issue: Medical devices - Sterility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
123364
123851
Synchro Neuro GuidewireM00313010
M00316310
136915
140598
136914
119461
140512
136913
Trevo Trak 21 Microcatheter90338
130285
129882
Trevo Pro 14 Microcatheter90231

Issue

Stryker Neurovascular has distributed multiple lots of Trevo Microcatheters and Synchro devices that did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with American National Standard Bacterial Endotoxins-Test methods, routine monitoring, and alternatives to batch testing AAMI ST72 standard. These lots of Stryker Neurovascular devices may exceed the acceptable limits of endotoxin levels.

Recall Start Date: April 29, 2024