· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| 123364 123851 | Synchro Neuro Guidewire | M00313010 M00316310 |
| 136915 140598 136914 119461 140512 136913 | Trevo Trak 21 Microcatheter | 90338 |
| 130285 129882 | Trevo Pro 14 Microcatheter | 90231 |
Stryker Neurovascular has distributed multiple lots of Trevo Microcatheters and Synchro devices that did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with American National Standard Bacterial Endotoxins-Test methods, routine monitoring, and alternatives to batch testing AAMI ST72 standard. These lots of Stryker Neurovascular devices may exceed the acceptable limits of endotoxin levels.
Recall Start Date: April 29, 2024