Health Products

Wavewriter Alpha Implantable Pulse Generator Kit

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Wavewriter Alpha Implantable Pulse Generator Kit
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Wavewriter Alpha Implantable Pulse Generator KitSC-1232
More than 10 numbers, contact manufacturer.Wavewriter Alpha 16 Implantable Pulse Generator KitSC-1216

Issue

Boston Scientific has received reports of undesired sensation and/or awareness of IPG turning off and on during IPG charging for patients with rechargeable SCS IPGs. Investigation of these reports determined that the device may reset (by design) due to detection of noise or interference during IPG charging. When the system resets, the patient's programmed stimulation turns off for approximately 10-15 seconds and then turns back on. When this occurs, some patients may perceive the sudden changes in stimulation. As soon as the reset is complete, the IPG resumes normal operation, including delivery of stimulation therapy. Boston Scientific has identified improvements (including an IPG firmware update for implanted devices) that will prevent this device reset behavior from occurring during IPG charging. Updating the IPG firmware eliminates the possibility of a coincident routine system check during IPG charging, thus preventing a potential system reset. This firmware update is available for any rechargeable SCS IPGs identified.

Recall Start Date: July 17, 2024