Health Products

Guidewire 70cm X 1mm (0.038 Inch)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: Guidewire 70cm X 1mm (0.038 Inch)
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Guidewire 70cm X 1mm (0.038 Inch)5583705

Issue

BD has received multiple reports that guidewire code 5583705, which is intended for use with the Bard 14.5 Fr Long-Term Dual Lumen Hemodialysis Catheters, may have been misassembled within the hoop such that the proximal (stiff) end of the guidewire is located at the distal end of the hoop. When loaded correctly the distal (flexible) end of the guidewire should be located at the distal end of the guidewire hoop.

Recall Start Date: August 1, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime