· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| Contact the manufacturer. | Tru-Core™ II | 763116160X 763120100X 763118100X 763418200X 763418250X 763114200X 763118200X 763114100X 763120160X 763116100X |
| Contact the manufacturer. | BioPince™ | 360-1580-01 360-1080-01 370-1580-01 360-2080-01 370-1080-01 |
| Contact the manufacturer. | BioPince™ Ultra Full Core Biopsy Instrument | 370-1580-02 360-1080-02 370-1080-02 360-1580-02 360-2080-02 |
| Contact the manufacturer. | BioPince™ Ultra Full Cpre Biopsy Instrument With Co-Axial Introducer Needle | 360-1080-03 360-2080-03 370-1580-03 360-1580-03 370-1080-03 |
A complaint was received regarding a small hole in the sterile barrier tray that is utilized for three products, BioPince™, BioPince™ Ultra and Tru-Core™ II automatic biopsy instruments. Inspection of inventory identified the presence of the defect in ~0.29% of product manufactured in the period.
Recall start date: July 3, 2024