· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| 140 | ADVIA® Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP) Assay | 11319151 11097543 |
| 130414 | Atellica® CH Urinary/Cerebrospinal Fluid Protein (UCFP) | Not applicable. |
Siemens Healthineers has confirmed the potential for biased quality control (qc) and patient results when using Atellica® CH Urinary/Cerebrospinal Fluid Protein (UCFP) lot 130414 on the Stellica® CH and Stellica® CI analyzers. Siemens' internal investigation confirmed when using lot 130414, QC may recover outside of the allowable control limits for urine chemistry and spinal fluid control levels.
Recall start date: August 1, 2024