Health Products

ADVIA® and Atellica® Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP) Assay

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology, Health products - Medical devices - Chemistry
Product: ADVIA® and Atellica® Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP) Assay
Issue: Medical devices - Device compatibility
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
140ADVIA® Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP) Assay11319151
11097543
130414Atellica® CH Urinary/Cerebrospinal Fluid Protein (UCFP)Not applicable.

Issue

Siemens Healthineers has confirmed the potential for biased quality control (qc) and patient results when using Atellica® CH Urinary/Cerebrospinal Fluid Protein (UCFP) lot 130414 on the Stellica® CH and Stellica® CI analyzers. Siemens' internal investigation confirmed when using lot 130414, QC may recover outside of the allowable control limits for urine chemistry and spinal fluid control levels.

Recall start date: August 1, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime