Dimension Vista® System - Total Bilirubin Assay

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Dimension Vista® System - Total Bilirubin Assay
Issue: Medical devices - Device compatibility
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
23206BADimension Vista® System - Total Bilirubin AssayK1167

Issue

Siemens Healthineers has confirmed, through investigation of customer complaints, the potential for discordant total bilirubin (tbil) results when the last three tests from the 160 test flex of lot 23206bA (wells 8 or 10) were processed on the Dimension Vista® platform. Calibrator, quality control and patient results could be impacted. The impact is variable across the analytical measurement range. This issue is isolated only to a portion of the flex lot therefore, the probability of occurrence of a discordant tbil result was determined to be 1%. Results could be erroneously increased or depressed due to the uncertain delivery of reagents from underfilled wells into the reaction cuvette.

Recall start date: August 6, 2024