Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Azurion 7 M20
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
Not applicable.722 224
722 079
Azurion 7 M20

Issue

Philips has identified that grease leaked from the bearings and/or anti-corrosion oil applied on the bearings can result in excessive lubrication on the ceiling rail. This can reduce the friction between the friction wheel and the rails. The reduced traction force causes the friction wheel to slip on the rail, leading to a "position slip" error mode being detected by the software. When this error occurs, the x-y-rz joystick is unable to send motorized requests to the relevant axes. Due to this issue, the motorized longitudinal movement of the Flexarm stand may be inconsistent (not smooth) and eventually may become unavailable.

Recall start date: August 7, 2024