BD Facsduet™ Sample Preparation Systems

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: BD Facsduet™ Sample Preparation Systems
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
9220010275
9220010313
662588Bd Facsduet™ Sample Preparation System (Base)
9230000104
9230000114
9230000112
9230000105
9230000138
9230000106
666339Bd Facsduet™ Premium Sample Preparation System

Issue

Through a customer complaint investigation, BD identified a specimen volume bias issue with the BD Facsduet™ sample preparation system when using software version 1.4.1. When the multidispense feature is enabled in the BD Facsduet™ sample preparation system software, and the user is preparing a two-tube assay, the specimen volume dispensed into the first tube will be at least 3% higher than the specimen volume dispensed into the second tube. For assays, where absolute counts are calculated using BD Trucount™ tubes, an increase in the volume may result in higher-than-expected absolute counts in the first tube compared to the second tube. This issue will be observed when preparing the following assays on the BD Facsduet™ sample preparation system: 1. BD Multitest™ IMK kit with BD Trucount™ tubes. note: if using the BD Multitest™ IMK kit with BD Trucount™ tubes, a QC message will appear on your report if the bias is outside the acceptable range per assay requirements. 2. BD Multitest™ 6-color TBNK with BD Trucount™ tubes only if consecutive replicates of two are run. 3. for all other quantitative assays utilizing BD Trucount™ tubes in consecutive replicates of two, please follow your institution's procedures to determine if a review of reported results is necessary. This issue does not affect assays that are determining percent positive, semi quantitative information.

Recall start date: August 12, 2024