Health Products

BioFire® FilmArray® 2.0 Instruments and BioFire® FilmArray® Torch Module Box

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: BioFire® FilmArray® 2.0 Instruments and BioFire® FilmArray® Torch Module Box
Issue: Medical devices - Performance
What to do: Contact the importer, bioMérieux Canada.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.BioFire® FilmArray® 2.0 InstrumentsFLM2-ASY-0001
More than 10 numbers, contact manufacturer.BioFire® FilmArray® Torch Module BoxHTFA-SUB-0103

Issue

Biomérieux has identified an issue with a subset of internal equipment used during the temperature calibration process on BioFire torch modules and 2.0 instruments manufactured or serviced since May 2022. The investigation found that BioFire instruments calibrated with this equipment may report higher melting temperatures (TMS). The impacted instruments must be returned to a BioFire service center for re-calibration.

Recall Start Date: August 9, 2024