Health Products

F&P Airvo 3

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: F&P Airvo 3
Issue: Medical devices - Performance
What to do: Customer will be contacted to arrange for software update.
Audience: General public, Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Contact the manufacturer.F&P Airvo 3PT301CA

Issue

This recall relates to specific versions of software of the Airvo 3, and how the device responds when flow alignment alarm 3.2.2 occurs.

When this happens in Airvo 3 devices set up with high pressure oxygen (HPO) and running software version 1.5.1 or earlier, the device will deliver room air only. If this happens, a patient may experience oxygen desaturation that could lead to hypoxia.

Software version 1.5.2 updates the software algorithm and ensures the target therapy continues in the event that this alarm occurs.