Health Products

Creatine Kinase (CK)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Creatine Kinase (CK)
Issue: Medical devices - Performance
What to do: Contact the ROCHE CARE CENTER at 1-877-273-3433 or (450) 686-7111 and by e-mail at LAVAL.TECHINFO@ROCHE.COM.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
755471Creatine Kinase (CK)07190794190

Issue

Roche received complaints from US customers for Creatine Kinase (CK) lot 755471 (expired June-2024) on cobas c 311/501/502 and INTEGRA 400 plus about abnormal calibration signals, flagged calibrations (Sens.E) and control recovery issues. Subsequent investigation identified a positive bias at the lower measuring range and a negative bias at the higher measuring range. The issue affects a small number of cassettes of the mentioned lot.

Recall Start Date: August 19, 2024