Health Products

Centricity PACS System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Centricity PACS System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Centricity PAC V3 - ALL
Centricity PAC V6 - ALL
Centricity PAC V2 - ALL
Centricity PAC V4 - ALL
Centricity PAC V7 - ALL
Centricity PACS SystemK2033JT
CENTRICITY PACS 3.1.1.1
00840682104807
K2024W
00840682145572
K2000YW

Issue

GE Healthcare has become aware of an issue where image storage is not possible when Centricity PACS reaches the maximum number of unique image identifiers (Ckeys). If this occurs, it can result in acquisition of a partial study. In an unlikely scenario that the partial study is unnoticed, it can lead to misdiagnosis. Centricity PACS is able to uniquely identify approximately 2.1 billion images (e.g. equivalent to 21 million exams with 100 images each). However, there is no notification to the user when the maximum number of unique image identifiers has been reached.

Recall Start Date: August 22, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime