Health Products

BD Alaris PCU

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: BD Alaris PCU
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
More than 10 numbers, contact manufacturer.8015BD Alaris PCU

Issue

BD has identified an increase in bent contacts on new female/left IUI connectors on devices and spare parts manufactured between Jan 01, 2023 and June 06, 2024. BD was made aware of this issue through complaints identified during the device check-in process and has not received any reports of this issue occurring during patient use. A defective pin on certain female/left IUI connectors that are not detected prior to use could result in an inoperable device or an operable device where adjoining device cannot be powered on at start up.

Recall start date: August 16, 2024