Ingenia 3.0T

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Ingenia 3.0T
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
s/n: 34055782103Ingenia 3.0T

Issue

Philips has identified an issue in the MR system patient support where the integrated radio frequency (IRF) carrier assembly screws may come loose and protrude beyond the surface. This could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop movement. The IRF carrier, a component of the patient support within the MR system, supports the tabletop where the patient lies. As the patient moves into the MRI tunnel, the tabletop, supported by the IRF carrier, forms a container for positioning the posterior coil.

Recall start date: September 4, 2024