IntelliVue MX550 Patient Monitor

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: IntelliVue MX550 Patient Monitor
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Not applicable.IntelliVue MX550 Patient Monitor866066

Issue

Philips became aware of one IntelliVue power supply with a broken ground bolt upon disconnection of the ground cable from the equipotential ground connector. Loss of electrical grounding may negatively affect the device's electromagnetic immunity and emission. Degraded electromagnetic immunity can cause the monitor to generate unreadable or unusable waveforms, potentially leading to incorrect/delayed patient treatment. Excessive or unintended electromagnetic emissions can also negatively impact the function of equipment in the vicinity of the patient monitor, potentially leading to delayed procedure. Although unlikely, these scenarios could potentially result in patient harm.

Recall Start Date: September 4, 2024