Health Products

MEERA Operating Tables

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Physical medicine
Product: MEERA Operating Tables
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
1086
600
1070
1075
1083
582
1084
1091
1085
MEERA Operating Tables720001F0
720001F2

Issue

It is possible that the error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control. With a cable-hand control the error does not occur. If the error "50037" occurs the movement is not permanently blocked.

Recall Start Date: September 11, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime