· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| 2005278 2006358 | Ultrasound System Bkactiv | 2300 |
| More than 10 numbers, contact manufacturer. | Bk3000/3500 Ultrasound Scanner System | 2300 |
| More than 10 numbers, contact manufacturer. | Bk5000 Ultrasound System | 2300 |
BK medical, a GE Healthcare Company, has become aware of an issue where the internal power connections at the power inlet of the device can come loose for bk3500, bk3000, bk5000 & bkactiv 2300 ultrasound devices with the battery option. In the unlikely event that this goes unnoticed, there is the potential for excessive heat, which could lead to a thermal event or the device becoming inoperable. There have been no injuries reported as a result of this issue.
Recall Start Date: September 11, 2024