Health Products

Bk3000/3500, Bk5000 and Bkactiv Ultrasound Systems

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology, Health products - Medical devices - Anaesthesiology
Product: Bk3000/3500, Bk5000 and Bkactiv Ultrasound Systems
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
2005278
2006358
Ultrasound System Bkactiv2300
More than 10 numbers, contact manufacturer.Bk3000/3500 Ultrasound Scanner System2300
More than 10 numbers, contact manufacturer.Bk5000 Ultrasound System2300

Issue

BK medical, a GE Healthcare Company, has become aware of an issue where the internal power connections at the power inlet of the device can come loose for bk3500, bk3000, bk5000 & bkactiv 2300 ultrasound devices with the battery option. In the unlikely event that this goes unnoticed, there is the potential for excessive heat, which could lead to a thermal event or the device becoming inoperable. There have been no injuries reported as a result of this issue.

Recall Start Date: September 11, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime