Health Products

SOLTIVE™ Premium SuperPulsed Laser System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: SOLTIVE™ Premium SuperPulsed Laser System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.SOLTIVE™ Premium SuperPulsed Laser SystemTFL-PLS

Issue

Olympus received a complaint regarding the SOLTIVE™ Laser System preset laser settings. The term "bladder stone" was incorrectly translated in both Spanish and Portuguese to "kidney stone" (cálculo renal) on the systems' graphical user interface (GUI). As a result of the incorrect Spanish and Portuguese translation from "bladder stone" to "kidney stone" on the GUI, there is a potential to deliver an incorrect or unintended amount of energy to the patient's anatomy. This issue is specific to the preset laser settings and does not affect users who are not utilizing the preset option.

Recall start date: September 12, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime