Health Products

EZDilate™ Balloon Dilator (WG)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Ear, nose and throat
Product: EZDilate™ Balloon Dilator (WG)
Issue: Medical devices - Labelling and packaging
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
408987EZDilate™ Balloon Dilator (WG)BD-410X-1055

Issue

Olympus is taking this recall action after having received 4 complaints between March 2024 and June 2024 from customers indicating the balloon was packaged with the incorrect Glo cath label applied to the balloon dilation catheter for the diameter sizes. The incorrect label states 11 mm, 12 mm, 13 mm however the balloon is 8.5mm, 9.5mm and 10.5 mm. All other labeling is correct for the balloon, including the shelf box with front box label and circular star label and the balloon pouch.

Recall start date: September 13, 2024