· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| Version # 4.20,4.30,5.10,5.20,5.30 | BD Synapsys™ Informatics Solution | 444150 |
BD identified through one (1) complaint that for BD Kiestra™ customers with BD Synapsys™, running Synapsys™ software versions 4.20, 4.30, 5.10, 5.20, or 5.30 and utilizing the in-line breakpoint feature, an error is occurring where BD Synapsys™ will incorrectly display breakpoints for organisms that do not have defined breakpoints in CLSI or EUCAST. This occurs when the identified organism is included in the same interpretive group as a species for which there is a specific breakpoint.
Recall start date: September 12, 2024