Health Products

Chameleon™ PTA Balloon Catheter

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Chameleon™ PTA Balloon Catheter
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Chameleon™ PTA Balloon CatheterCH07-40-75US
CH12-40-75US
CH08-40-75US
CH06-40-75US
CH05-40-75US
CH10-40-75US

Issue

Specific lots of the device failed to meet in-house balloon pressure testing. There were no serious patient injuries or deaths reported. While not observed or reported, additional potential risks or harms may include delay to treatment, hemorrhaging, blood loss, bleeding, or perforation of vessels. The most foreseeable sequence of events would be balloon rupture during inflation while under close monitoring and traditionally direct fluoroscopic visualization, resulting in careful removal of the malfunctioned device and replacement with a new balloon catheter for procedure completion.

Recall start date: September 19, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime