Health Products

Sterilizable Internal Defibrillation Paddles

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Sterilizable Internal Defibrillation Paddles
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
106917875
96496376
107445069
107078533
107556560
107445059
11131-000045
11131-000047
11131-000042
Sterilizable Internal Defibrillation Paddles
9964902511131-000044Sterilizable Internal Defibrillation Paddles
107124605
101506304
11131-000043
11131-000046
Sterilizable Internal Defibrillation Paddles

Issue

Stryker has received complaints indicating issues with the internal paddles. Complaints which state that the paddles are not reaching the expected cleaning and sterilization cycle count designated in the instructions for use and are showing signs of cracking. A CAPA was opened November 2023.

Recall start date: September 20, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime