Health Products

Olympus Flushing Pump

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Gastroenterology and urology
Product: Olympus Flushing Pump
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
More than 10 numbers, contact manufacturer.OFP-2Olympus Flushing Pump

Issue

During product testing, Olympus identified an intermittent loss of function of the ofp-2 flushing pump caused by an internal component connection failure. in the event that the connection failure occurs, the ofp-2 device will return to or remain in the "off" condition and the operator will not be able to use the flushing function. Through its investigation, Olympus determined that this issue resulted from an alternate tool being used during the manufacturing of one (1) lot of the subject component. Olympus has received three (3) complaints that are potentially applicable to this issue. No adverse events have been reported.

Recall start date: September 20, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime