Health Products

Disposable Pressure Transducers

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Disposable Pressure Transducers
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
More than 10 numbers, contact manufacturer.MX9502T
MX9505T
MXG300231
MX9504T
MX9510T
MX950L
MX950
MX9501T
Disposable Pressure Transducers

Issue

Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the disposable pressure transducer or from a defect within the pressure transducer chip.

Recall start date: September 20, 2024