Health Products

Ultraview DM3 Monitor

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Ultraview DM3 Monitor
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Ultraview DM3 Monitor91330

Issue

The lithium-ion battery used in the DM3 is susceptible to failure and overheating when used beyond its recommended lifetime or otherwise damaged. This is due to a short in the battery with indeterminant cause, but likely attributed to use beyond service life, overcharging, or damage to the battery. Batteries known to be older than two years or batteries that are no longer holding a charge should be replaced. Damaged batteries must be replaced immediately.

Recall Start Date: September 27, 2024

Get told about the next health product recall.
Get health product recall alerts