Health Products

Zenition 70 and 50

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Zenition 70 and 50
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Not applicable.Zenition 70718133
Not applicable.Zenition 50718096

Issue

During the system start up or during an ongoing clinical procedure, due to fluctuations in the hospital main power, the fuse can blow thereby causing loss of power to the generator and height positioning module. This results in a system not being available for clinical use. As the system is not available for clinical use, it can result in a delay or abortion of the clinical procedure.

Recall Start Date: September 30, 2024

Get told about the next health product recall.
Get health product recall alerts