· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Uropass Ureteral Access Sheath | 61038BX 61054BX 61146BX 61246BX 61138BX 61324BX 61024BX 61046BX 61254BX 61154BX 61346BX 61354BX 61124BX 61238BX 61224BX 61338BX | Not applicable. |
Olympus conducted an investigation after receiving complaints reporting broken dilator tips in the package and in patients during surgical procedures. The investigation determined that exposing the Uropass product to ultraviolet ("UV") radiation can cause brittleness of the device dilator tip, which may lead to breakage. Since April 2023, Olympus has received 2 adverse event complaints reporting broken Uropass dilator tips for devices still within their shelf life.
Recall start date: October 4, 2024