Health Products

TDHistoCyto Software

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: TDHistoCyto Software
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
TDHistoCyto SoftwareVersion 13.61 and laterTD-TDH-01

Issue

A software bug was reported by a user site. The end user noticed a discrepancy between the values of a number of additional data displayed in MS Word format compared with what is printed, including what is displayed in the PDF report after validation. When the problem occurs, the PDF report that can be consulted by prescribers does not include the latest updated values of additional data.

Recall start date: October 15, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime