Health Products

THS™ Neonatal Cart With Powervar Security II Medical UPM

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: THS™ Neonatal Cart With Powervar Security II Medical UPM
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
THS™ Neonatal Cart With Powervar Security II Medical UPM5022072R-2330003
5022072R-2330006
5022072R-2330005
5022072R-2330001
5022072R-2330004
5022072R-2330002
50220-72R

Issue

Dräger has been made aware of an incident involving the UPS battery on the Technimount THS™ Neonatal Cart switching off during patient transfer with the Dräger incubator Babyleo® TN500. There were no injuries or impact on patient care resulting from this incident, however Dräger Medical Canada Inc., the distributor of the cart has deemed necessary to initiate a recall to advise users to adhere to the battery IFU to prevent a similar failure.

Recall start date: October 1, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime