Health Products

RENASYS™ EDGE Negative Pressure Wound Therapy

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: RENASYS™ EDGE Negative Pressure Wound Therapy
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
RENASYS™ EDGE Negative Pressure Wound TherapyMore than 10 numbers, contact manufacturer.66803126

Issue

Smith & Nephew Medical Ltd., has initiated a field action to voluntarily correct RENASYS™ EDGE devices. If the batteries are not appropriately maintained and charged as defined in the IFU (left in storage in an uncharged state), the battery may become depleted and be unable to recharge. The RENASYS™ EDGE device will operate on mains power and deliver therapy as intended. The device will show the charger is connected but will not increase battery charge percentage.

Recall start date: October 2, 2024