Health Products

Acrobat-i Stabilizer Z and Acrobat-i Positioner Z

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Acrobat-i Stabilizer Z and Acrobat-i Positioner Z
Issue: Medical devices - Sterility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
Acrobat-i Stabilizer ZMore than 10 numbers, contact manufacturer.OM-10000Z
Acrobat-i Positioner Z3000290368
3000353846
3000308238
3000334128
3000355540
XP-5000Z

Issue

The contract manufacturer of the Acrobat product family reported deviations in the manufacturing process that creates the sterile barrier. Specifically, in some instances, the tray sealing step was performed multiple times using the same package, which is outside of the validated process. Although testing suggests that product is sterile immediately after being sealed with this non-validated process, we cannot ensure that the product will remain sterile for the duration of its 2-year shelf life.

Recall Start Date: October 21, 2024