· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| JuggerLoc Disposable Cannula Kit - Drill Guide | 0002467342 0002476076 0002497715 0002504035 0002446333 0002493107 0002496066 0002446326 | 110027357 |
Zimmer Biomet is conducting a lot specific medical device recall for the JuggerLoc Disposable Cannula Kit - Drill Guide. Zimmer Biomet has received multiple complaints where it was reported that the guides were discernibly warped when pulled out of packaging. Following investigation, the warping has been attributed to the cleaning process used during supplier manufacturing.
Recall Start Date: October 28, 2024