Health Products

TCM5 monitor

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: TCM5 monitor
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Tcm5software versions 1.5.0
software versions 1.5.0.1
393-500

Issue

Radiometer has become aware of an issue with the TCM5 monitor. The issue is that the TCM5 application may freeze and not update the display with new values or send new values to a central monitoring system. This means that the last measured values remain displayed unchanged on the screen of the TCM5 screen and in a central monitoring system and may not reflect patient status. For this reason, Radiometer requests that you stop using the affected product (except in sleep studies*), with immediate effect. *All TCM5 monitors with software versions 1.5.0 or 1.5.0.1 are affected. However, it has been assessed that if used in sleep studies, non-critical patients will not experience harm.

Recall start date: October 22, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime