Health Products

BIORAPTOR™ Suture Anchors

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Orthopaedics
Product: BIORAPTOR™ Suture Anchors
Issue: Medical devices - Sterility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Bioraptor Curved 2.3 Pk Suture Anchor2151692
2151422
72203281
72203280
Bioraptor 2.3 Pk Suture Anchor W/Ultrabraid Suture2151690
2151242
2146369
2151691
2153489
72201542
72201541

Issue

Smith & Nephew, Inc. has initiated a field action to voluntarily remove seven (7) lots of BIORAPTOR™ Suture Anchors due to a potential for sterile barrier breach. Under certain shipping conditions, the distal end of the inserter can potentially pierce the packaging tray, compromising the sterile barrier.